For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Lactoglobulin tryptic map on Luna C18(2) - 0.0125% TFA
14202
Separation Mode: Reversed Phase
Reversed Phase
C18(2)
HPLC
Peptide
Biochemical Compound / Nutrient

Applications

Lactoglobulin tryptic map on Luna C18(2) - 0.0125% TFA
Analytes
1Protein

Analyte not found

Details
LC Conditions (App ID: 14202)
Column:
Brand Name:
Part No:
Phase Name:
C18(2)
Elution Type:
Gradient
Mobile Phase:
A: Water with 0.0125% TFA
B: Acetonitrile with 0.0125% TFA
Gradient Profile:
Step No.Time(min)%A%B
10973
23973
3906040
Flow Rate:
1 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 210 nm (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Similar Applications
search
No Data