Region

Applications

Extraction of Acetaminophen on Strata™ X

Sample Preparation Details

Condition 1 :1.0 mL Methanol
Condition 2 :1.0 mL Water
Load :Load 1.00 mL Sample diluted in Buffer
Wash 1 :1.0 mL 5% methanol/D. I. water
Dry :For 1 min at 10'' of Hg
Elution 1 :1.0 mL Methanol
Evaporate: :Evaporate to dryness
Reconstitute :2 mL Mobile Phase
Notes :Blow down to almost dryness under slow stream of N2. Reconstitute with mobile phase to 200 uL. Inject 50 uL on HPLC

Application Detail (App ID: 14452)

Part No:00F-4249-E0
Dimensions:150 x 4.6 mm ID
Elution Type:Isocratic
Elution A:20 mM KH2PO4 (pH 7.0)
Elution B:Methanol
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 90 10

Flow Rate:1 mL/min
Col. Temp: ambient
Detection: UV-Vis Abs.-Fixed Wave. (UV) @ 220 nm (25°C)
Note:SecurityGuard™ Guard Cartridge System extends column lifetime.
- SecurityGuard Cartridges, C8 4 x 3.0mm, 10/Pk Part No.: AJ0-4290
- Holder Part No.: KJ0-4282

Analyte Details

Analyte Detail not available

1
Sulfanilimide
search

2
Acetaminophen
search