FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Anilines - Luna 5µm C18(2)
15692
Separation Mode: Reversed Phase
Reversed Phase
C18(2)
HPLC
Carcinogens
Organic Chemical
Chemical Compound

Applications

Anilines - Luna 5µm C18(2)
Analytes
1Caffeine
2Aniline
3N,N-Dimethylaniline
42-Nitroaniline
5N-Methylaniline
62-Ethylaniline
Compound Name
Caffeine
CID
2519
Molecular Formula
C8H10N4O2
Molecular Weight
194
No. Hydrogen Bond Acceptors
3
Details
LC Conditions (App ID: 15692)
Column:
Brand Name:
Part No:
Phase Name:
C18(2)
Elution Type:
Gradient
Mobile Phase:
A: 10mM NH4HCO3, pH 10.5
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
108020
2105050
Flow Rate:
1 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 254 nm (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282

Order Items Used in This Application