FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Acyclovir and Gancylovir analysis using Synergi Fusion-RP
15741
Separation Mode: Reversed Phase
Reversed Phase
Fusion-RP
HPLC
Antiviral Agents
Pharmaceutical
Therapeutic / Clinical

Applications

Acyclovir and Gancylovir analysis using Synergi Fusion-RP
Analytes
1Acyclovir
2Ganciclovir
Compound Name
Acyclovir
CID
2022
Molecular Formula
C8H11N5O3
Molecular Weight
225
No. Hydrogen Bond Acceptors
6
No. Hydrogen Bond Donors
3
Details
LC Conditions (App ID: 15741)
Column:
Brand Name:
Part No:
Phase Name:
Fusion-RP
Elution Type:
Isocratic
Mobile Phase:
A: 20mM Ammonium Formate, pH 3.5
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
10973
Flow Rate:
1 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 260 nm (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Fusion-RP 4 x 3.0mm, 10/Pk Part No.: AJ0-7557 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Similar Applications
search
No Data