FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
MercuryMS - Luna 3µm C18(2) 20 x 2.0mm for Benzodiazepines in Plasma by LC/MS/MS
16040
Separation Mode: Reversed Phase
Reversed Phase
C18(2)
HPLC
Antiemetics
Clinical
Forensics/Toxicology
Toxicology / Drug Screening / TDM
Therapeutic / Clinical

Applications

MercuryMS - Luna 3µm C18(2) 20 x 2.0mm for Benzodiazepines in Plasma by LC/MS/MS
Analytes
17-Aminoflunitrazepam
2Chlordiazepoxide
3Oxazepam
4Nordazepam
5Clonazepam
6Temazepam
7Flunitrazepam
8Diazepam
Compound Name
7-Aminoflunitrazepam
CID
92294
Molecular Formula
C16H14FN3O
Molecular Weight
283
No. Hydrogen Bond Acceptors
4
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 16040)
Column:
Brand Name:
Part No:
Phase Name:
C18(2)
Elution Type:
Gradient
Mobile Phase:
A: 0.1% Formic Acid in Water
B: 0.1% Formic Acid in Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
108515
21595
31.018515
428515
Flow Rate:
0.6 mL/min
Temperature:
22°C
Detection:
Mass Spectrometer (MS) (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C8 4 x 2.0mm, 10/Pk Part No.: AJ0-4289 - Holder Part No.: KJ0-4282