For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Nicotine and Metabolites on Gemini 5µm C18
16345
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Indicators and Reagents
Clinical
Pharmaceutical
Therapeutic / Clinical

Applications

Nicotine and Metabolites on Gemini 5µm C18
Analytes
1Nornicotine
2Nicotine
3Cotinine
Compound Name
Nornicotine
CID
412
Molecular Formula
C9H12N2
Molecular Weight
148
No. Chiral Centers:
1
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 16345)
Column:
Brand Name:
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: Water with 0.1% v/v Formic Acid
B: Acetonitrile with 0.1% v/v Formic Acid
Gradient Profile:
Step No.Time(min)%A%B
10973
21973
337030
43.01973
55.5973
Flow Rate:
0.4 mL/min
Temperature:
22°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
SCIEX API 3000™ Triple Quadrupole Mass Spectrometer with TurboIonSpray Source MRM transitions monitored: Nicotine: m/z 163.3/105.9 and 163.3/80.1 Nornicotine: m/z 149.1/80.1 Cotinine: m/z 177.3/98.0 and 177.3/80.1
Sample Note:
The pH of the aqueous acid is 2.7. 0.5ng on-column injection of each analyte.

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm EVO C18 100 Å, LC Column 100 x 3.0 mm, EA
Reversed Phase
Kinetex 5 µm EVO C18 100 Å, LC Column 100 x 3.0 mm, Ea
Reversed Phase
Gemini 3 µm NX-C18 110 Å, LC Column 100 x 2 mm, Ea
Reversed Phase
Gemini 3 µm NX-C18 110 Å, LC Column 50 x 2 mm, Ea
Reversed Phase
Luna 5 µm HILIC 200 Å, LC Column 100 x 2 mm, Ea
Hydrophilic Interaction