For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
USP 467 Procedure B - Class 1 by GC/FID on ZB-WAXplus
16352
Separation Mode: Improved Performance
Improved Performance
USP <467>
ZB-WAXplus
GC
Pharmaceutical/Biopharmaceutical

Applications

USP 467 Procedure B - Class 1 by GC/FID on ZB-WAXplus
Analytes
11,1-Dichloroethene
2Carbon tetrachloride
31,1,1-Trichloroethane
4Benzene
51,2-Dichloroethane
Compound Name
Vinylidene Chloride
CID
6366
Molecular Formula
C2H2Cl2
Molecular Weight
96
Details
Official Method: USP <467>
GC Conditions (App ID: 16352)
Column:
Part No:
Phase Name:
ZB-WAXplus
Injection:
Headspace 5:1 @ 140°C, 1.0 μL
Oven Program:
50 °C for 20 min to 165 °C @ 6 °C/min for 20 min
Carrier Gas:
Constant Flow Helium, 35.0 cm/sec
Detection:
Flame Ionization (FID) (250°C)
Sample Notes:
Sample prepared as per USP Method

Order Items Used in This Application

Similar Applications
search
Zebron ZB-624, GC Cap. Column 30 m x 0.32 mm x 1.80 µm, Ea
Standard
Zebron ZB-WAXplus, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Improved Performance