For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA Method 8330: Explosive Mix Using Synergi Hydro-RP
18146
Separation Mode: Reversed Phase
Reversed Phase
Hydro-RP
HPLC
Vasodilator Agents
Environmental
Pollutant / Contaminant
Environmental / Health

Applications

EPA Method 8330: Explosive Mix Using Synergi Hydro-RP
Analytes
1Octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazocine (HMX)
2Hexahydro-1,3,5-trinitro-1,3,5-triazine
3Picric acid
41,3,5-Trinitrobenzene (1,3,5-TNB)
5Methyl-2,4,6-trinitrophenylnitramine
61,3-Dinitrobenzene (1,3-DNB)
7Nitrobenzene
8Nitroglycerin
92,4,6-Trinitrotoluene (2,4,6-TNT)
103,5-Dinitroaniline
114-Amino-2,6-dinitrotoluene
122-Amino-4,6-dinitrotoluene
132,6-Dinitrotoluene (2,6-DNT)
142,4-Dinitrotoluene (2,4-DNT)
15Pentaerythritol tetranitrate (PETN)
162-Nitrotoluene
174-Nitrotoluene
183-Nitrotoluene
Compound Name
1,3,5,7-Tetrazocane
CID
80156
Molecular Formula
C4H12N4
Molecular Weight
116
No. Hydrogen Bond Acceptors
4
No. Hydrogen Bond Donors
4
Details
LC Conditions (App ID: 18146)
Column:
Brand Name:
Part No:
Phase Name:
Hydro-RP
Elution Type:
Gradient
Mobile Phase:
A: Methanol
B: 20mM KH2PO4-pH=2.5
Gradient Profile:
Step No.Time(min)%A%B
105545
225545
3126535
4226535
Flow Rate:
1 mL/min
Temperature:
25°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 254 nm (22°C)
Sample Note:
100ul of 100ug/ml 17 component mix was diluted with 900ul of 50:50 Methanol/water

Order Items Used in This Application

Similar Applications
search
Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Synergi 2.5 µm Hydro-RP 100 Å, LC Column 100 x 2 mm, Ea
Reversed Phase
Synergi 2.5 µm Polar-RP 100 Å, LC Column 100 x 2 mm, Ea
Reversed Phase
Synergi 4 µm Polar-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Gemini 5 µm NX-C18 110 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Gemini 5 µm C6-Phenyl 110 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Gemini 5 µm C18 110 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase