FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Aroclor 1260: DIN 51527 on ZB-35HT
18180
Separation Mode: High Temperature
High Temperature
ZB-35HT
GC
Pollutant / Contaminant
Environmental / Health

Applications

Aroclor 1260: DIN 51527 on ZB-35HT
Analytes
1TCMX
2BZ# 31
3BZ# 28
4BZ# 52
5BZ# 101
6BZ# 118
7BZ# 153
8BZ# 105
9BZ# 138
10BZ# 167
11BZ# 156
12BZ# 180
13BZ# 170
14BZ# 189
15DCB

Analyte not found

Details
GC Conditions (App ID: 18180)
Column:
Part No:
Phase Name:
ZB-35HT
Injection:
Splitless @ 250°C, 1.0 μL
Oven Program:
50 °C for 0.5 min to 220 °C @ 30 °C/min to 320 °C @ 6 °C/min for 2 min
Carrier Gas:
Constant Flow Helium, 0.5 mL/min
Detection:
Electron Capture (ECD) (350°C)
Detector Info:
Used a pressure pulse at 40 psi for 0.25 min
Sample Notes:
Total concentration of Aroclors was 90 ppm in iso-octane