Region

Applications

Cough and Cold Medicine on Kinetex 2.6µm C18 150 x 4.6mm ID

Application Detail (App ID: 18533)

Part No:00F-4462-E0
Dimensions:150 x 4.6 mm ID
Elution Type:Gradient
Elution A:0.1% phosphoric acid in water
Elution B:0.1% phosphoric acid in acetonitrile
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 95 5
2 0.75 95 5
3 6 50 50
4 6.75 10 90
5 6.76 95 5
6 9.75 95 5

Flow Rate:1.8 mL/min
Col. Temp: 50 °C
Detection: UV-Vis Abs.-Diode Array (PDA) @ 215 nm (22°C)
Note:SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime.
- SecurityGuard ULTRA Cartridges, UHPLC C18 for 4.6mm ID Columns, 3/Pk Part No.: AJ0-8768
- Holder Part No.: AJ0-9000

Analyte Details

Analyte Detail not available

1
impurities from Dextromethorphan

2
Maleate
search

3
Phenylephrine
search

4
Acetmainophen
search

5
Doxylamine
search

6
Chlorpheniramine
search

7
Dextromethorphan
search