Region

Applications

Ranitidine (Degraded) - Forced Degradation Stability by Kinetex 2.6µm C18 150 x 4.6mm ID

Application Detail (App ID: 18600)

Part No:00F-4462-E0
Dimensions:150 x 4.6 mm ID
Elution Type:Gradient
Elution A:0.1% formic acid in water
Elution B:0.1% formic acid in acetonitrile
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 95 5
2 7 80 20
3 9 5 95
4 9.01 95 5
5 14 95 5

Flow Rate:1.4 mL/min
Col. Temp: 30 °C
Detection: UV-Vis Abs.-Diode Array (PDA) @ 230 nm (22°C)
Note:SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime.
- SecurityGuard ULTRA Cartridges, UHPLC C18 for 4.6mm ID Columns, 3/Pk Part No.: AJ0-8768
- Holder Part No.: AJ0-9000

Analyte Details

Analyte Detail not available

1
Ranitidine
search

2
impurities