Region

Applications

EPA 531.1 Carbamate Pesticides at High Speed using Kinetex 2.6µm C18 50 x 2.1mm ID

Sample Preparation Details

Condition 1 :1 mL Methanol
Condition 2 :1 mL Loading buffer
Load :Load Sample
Wash 1 :1 mL Loading buffer
Wash 2 :1 mL Methanol/Water (15:85)
Dry :2 minutes at 10 mm Hg vacuum
Elution 1 :0.5 mL 5% Formic acid in Methanol
Elution 2 :0.5 mL 5% Formic acid in Methanol
Evaporate: :Evaporate to dryness
Reconstitute :1 mL Mobile phase

Application Detail (App ID: 18610)

Part No:00B-4462-AN
Dimensions:50 x 2.1 mm ID
Elution Type:Gradient
Elution A:Water with 0.1% H3PO4
Elution B:Acetonitrile with 0.1% H3PO4
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 95 5
2 3 5 95

Flow Rate:1 mL/min
Col. Temp: 40 °C
Detection: UV-Vis Abs.-Variable Wave.(UV) @ 210 nm (22°C)
Note:SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime.
- SecurityGuard ULTRA Cartridges, UHPLC C18 for 2.1mm ID Columns, 3/Pk Part No.: AJ0-8782
- Holder Part No.: AJ0-9000

Analyte Details

Analyte Detail not available

1
Aldicarb-sulfoxide
search

2
Aldicarb-sulfone
search

3
Oxamyl
search

4
Methomyl
search

5
3-Hydroxy-Carbofuran
search

6
Aldicarb sulfone-related impurity

7
Aldicarb
search

8
Baygon (Propoxur)
search

9
Carbofuran
search

10
Carbaryl
search

11
1-Napthol
search

12
Methiocarb
search