Region

Applications

LC/MS for Trypsin Digested BSA-peptide mapping on Kinetex 2.6 µm C18 150 x 2.1mm

Application Detail (App ID: 19031)

Part No:00F-4462-AN
Dimensions:150 x 2.1 mm ID
Elution Type:Gradient
Elution A:0.1% Formic Acid in water
Elution B:0.1% Formic Acid in Acetonitrile
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 97 3
2 40 64 36
3 42 5 95
4 42.1 97 3

Flow Rate:0.22 mL/min
Col. Temp: 25 °C
Detection: Mass Spectrometer (MS) @ amu (22°C)
Note:SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime.
- SecurityGuard ULTRA Cartridges, UHPLC C18 for 2.1mm ID Columns, 3/Pk Part No.: AJ0-8782
- Holder Part No.: AJ0-9000

Analyte Details

Analyte Detail not available

1
Bovine Serum Albumin (BSA)