Region

Applications

LC/MS for Trypsin Digested Myoglobin-peptide mapping on Kinetex 2.6 µm C18 150 x 4.6mm

Application Detail (App ID: 19034)

Part No:00F-4462-E0
Dimensions:150 x 4.6 mm ID
Elution Type:Gradient
Elution A:0.1% Formic Acid in water
Elution B:0.1% Formic Acid in Acetonitrile
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 97 3
2 25 32 68
3 27 5 95
4 27.1 97 3

Flow Rate:1.2 mL/min
Col. Temp: 25 °C
Detection: Mass Spectrometer (MS) @ amu (22°C)
Note:SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime.
- SecurityGuard ULTRA Cartridges, UHPLC C18 for 4.6mm ID Columns, 3/Pk Part No.: AJ0-8768
- Holder Part No.: AJ0-9000

Analyte Details

Analyte Detail not available

1
Myoglobin