BSA tryptic map in formic acid using Kinetex 2.6u C18 150x4.6mm

Application Detail (App ID: 19121)

Part No:00F-4462-E0
Dimensions:150 x 4.6 mm ID
Elution Type:Gradient
Elution A:Water with 0.1% formic acid
Elution B:Acetonitrile with 0.1% formic acid
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 97 3
2 30 50 50

Flow Rate:1 mL/min
Col. Temp: 35 °C
Detection: UV-Vis Abs.-Variable Wave.(UV) @ 214 nm (22°C)
Note:SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime.
- SecurityGuard ULTRA Cartridges, UHPLC C18 for 4.6mm ID Columns, 3/Pk Part No.: AJ0-8768
- Holder Part No.: AJ0-9000

Analyte Details

Analyte Detail not available

Bovine Serum Albumin (BSA)