FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Preservatives Analysis using Luna 3µm C18(2) 100x4.6mm
19500
Separation Mode: Reversed Phase
Reversed Phase
C18(2)
HPLC
Anti-Infective Agents, Local
Food & Beverage
Food and Beverage (incl. additives
contaminants)
Product / Trade Good

Applications

Preservatives Analysis using Luna 3µm C18(2) 100x4.6mm
Analytes
1Benzyl alcohol
2Phenoxyethanol
3Sorbic acid
4Benzoic acid
5Methyl paraben
6p-Anisic acid
7Dehydroacetic acid
8Salicylic acid
9Ethyl paraben
10Isopropyl paraben
11Propyl paraben
12Isobutyl paraben
13Butyl paraben
14Triclosan
15Triclocarban
Compound Name
Benzyl Alcohol
CID
244
Molecular Formula
C7H8O
Molecular Weight
108
No. Hydrogen Bond Acceptors
1
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 19500)
Column:
Brand Name:
Part No:
Phase Name:
C18(2)
Elution Type:
Gradient
Mobile Phase:
A: Water with 0.1% TFA
B: Acetonitrile with 0.1% TFA
Gradient Profile:
Step No.Time(min)%A%B
108515
2201585
Flow Rate:
1.5 mL/min
Temperature:
30°C
Detection:
UV-Vis Abs.-Diode Array (PDA) @ 214 nm (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm ID, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282