Propranolol by HPLC - Kinetex 2.6µm C8 100x4.6mm

Application Detail (App ID: 19522)

Part No:00D-4497-E0
Dimensions:100 x 4.6 mm ID
Elution Type:Isocratic
Elution A:0.5g SDS in 18mL of 150mM phosphoric acid , 90mL ACN, 90mL MeOH diluted to 250mL w/water
Gradient Profile:
Step No. Time (min) Pct A
1 10 100

Flow Rate:1.5 mL/min
Col. Temp: 25 °C
Detection: UV-Vis Abs.-Variable Wave.(UV) @ 290 nm (22°C)
Note:SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime.
- SecurityGuard ULTRA Cartridges, UHPLC C8 for 4.6mm ID Columns, 3/Pk Part No.: AJ0-8770
- Holder Part No.: AJ0-9000

Analyte Details

Analyte Detail not available