For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Quantification of Acidic Pesticides by QuEchERS and LC/MS/MS
20024
Separation Mode: Reversed Phase
Reversed Phase
Fusion-RP
HPLC
Hallucinogens
Food & Beverage

Applications

Quantification of Acidic Pesticides by QuEchERS and LC/MS/MS
Analytes
1Clopyralid
2Clopyralid
3Imazapyr
4Imazapyr
5Dicamba
6Dicamba
7Bentazon
8Bentazon
9Fluoxypyr
10Fluoxypyr
112,4,5-TP
122,4,5-TP
13Bromoxynil
14Bromoxynil
15MCPA
16MCPA
17Triclopyr
18Triclopyr
192,4-DB
202,4-DB
21Ioxynil
22Ioxynil
232,4,5–T
242,4,5–T
25MCPP
26MCPP
272,4-D
282,4-D
292,4–DP
302,4–DP
31MCPB
32MCPB
33Dinoterb
34Dinoterb
35Dinoseb
36PCP
37PCP
Compound Name
Lontrel
CID
15553
Molecular Formula
C6H3Cl2NO2
Molecular Weight
191
No. Hydrogen Bond Acceptors
3
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 20024)
Column:
Brand Name:
Part No:
Phase Name:
Fusion-RP
Elution Type:
Gradient
Mobile Phase:
A: 5mM Ammonium Formate in Water
B: 5mM Ammonium Formate in Methanol
Gradient Profile:
Step No.Time(min)%A%B
10955
21.56040
384060
4101090
5141090
615955
Flow Rate:
0.25 mL/min
Temperature:
25°C
Detection:
Tandem Mass Spec (MS-MS) (22°C)
Detector Info:
SCIEX API 3000™ or 3200 QTRAP
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Fusion-RP 4 x 2.0mm ID, 10/Pk Part No.: AJ0-7556 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Similar Applications
search
Synergi 2.5 µm Fusion-RP 100 Å, LC Column 50 x 2 mm, Ea