For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Aldosterone analysis by LCMS using Kinetex 2.6um XB-C18 50x3.0mm
20648
Separation Mode: Reversed Phase
Reversed Phase
XB-C18
HPLC
Clinical
Clinical Research
Therapeutic / Clinical

Applications

Aldosterone analysis by LCMS using Kinetex 2.6um XB-C18 50x3.0mm
Analytes
1Aldosterone
Compound Name
Aldosterone
CID
5839
Molecular Formula
C21H28O5
Molecular Weight
360
No. Hydrogen Bond Acceptors
5
No. Hydrogen Bond Donors
2
Details
LC Conditions (App ID: 20648)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
XB-C18
Elution Type:
Gradient
Mobile Phase:
A: Water
B: 50/50 AcCN/Methanol
Gradient Profile:
Step No.Time(min)%A%B
107030
241090
351090
45.17030
577030
Flow Rate:
0.5 mL/min
Temperature:
22°C
Detection:
Tandem Mass Spec (MS-MS) (22°C)
Sample Preparation Details
Description:
Strata-X-A 33 µm Polymeric Strong Anion, 60 mg / 3 mL, Tubes, 50/Pk
Part Number:
Condition 1:
Methanol
Condition 2:
25 mM Ammonium Bicarbonate
Wash 1:
25mm ammonium bicarbonate buffer
Wash 2:
25% methanol in water
Evaporation:
Evaporate to dryness
1
Condition
Methanol
2
Shake
Ethyl Acetate:Isopropanol:NH4OH (7:2:1)
1
Condition
Methanol
2
Shake
Ethyl Acetate:Isopropanol:NH4OH (7:2:1)
1
Condition
Methanol
2
Shake
Ethyl Acetate:Isopropanol:NH4OH (7:2:1)

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm XB-C18 100 Å, LC Column 50 x 3 mm, Ea