FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Histamine by LCMS using Luna 3um HILIC 100x2.1mm
20650
Histamine Agents
Clinical
Biochemical Compound / Nutrient
Therapeutic / Clinical

Applications

Histamine by LCMS using Luna 3um HILIC 100x2.1mm
Analytes
11-Methyl histamine
2N-Methyl histamine
3Histamine
Compound Name
1-Methylhistamine
CID
3614
Molecular Formula
C6H11N3
Molecular Weight
125
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 20650)
Elution Type:
Gradient
Mobile Phase:
A: 50:45:5 ACN:DI H2O: 100 mM Amm. Form, pH 3.22
B: 95:5 ACN:100 mM Amm. Formate, pH 3.22
Gradient Profile:
Step No.Time(min)%A%B
101783
21.51000
32.51000
42.61783
54.51783
Flow Rate:
0.6 mL/min
Temperature:
22°C
Detection:
Tandem Mass Spec (MS-MS) (22°C)
Detector Info:
SCIEX API 5000™
Sample Preparation Details
Description:
Strata-X-CW 33 µm Polymeric Weak Cation, 30 mg / 3 mL, Tubes, 50/Pk
Part Number:
Condition 1:
Methanol
Condition 2:
DI water
Wash 1:
DI water
Wash 2:
MeOH
Evaporation:
Evaporate to dryness
1
Condition
Methanol
2
Condition
DI water
1
Shake
87:13 B:A, fortified to contain 5% formic acid
1
Wash
DI water
2