For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Bath salts analysis using Kinetex 2.6u C18 50x3.0mm
20791
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Adrenergic Agents
Forensics/Toxicology

Applications

Bath salts analysis using Kinetex 2.6u C18 50x3.0mm
Analytes
1Methylone
2Methamphetamine
3Mephedrone
4MDPV

Analyte not found

Details
LC Conditions (App ID: 20791)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 10mM Ammonium formate
B: Methanol with 0.1% Formic acid
Gradient Profile:
Step No.Time(min)%A%B
10955
230100
340100
Flow Rate:
500 µL/min
Temperature:
22°C
Detection:
Tandem Mass Spec (MS-MS) (22°C)
Detector Info:
SCIEX API 4000™

Order Items Used in This Application