For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Cocaine and opiates by GCMS on a ZB-5MS
20809
Separation Mode: Standard
Standard
ZB-5MS
GC
Analgesics, Opioid
Forensics/Toxicology
Toxicology / Drug Screening / TDM
Therapeutic / Clinical

Applications

Cocaine and opiates by GCMS on a ZB-5MS
Analytes
1Ecgonine methyl ester
2Cocaine
3Cocaethylene
4Benzoylecgonine
5Dihydrocodeine
6Codeine
7Ethylmorphine
8Morphine
96-MAM
Compound Name
Methyl Ecgonine
CID
104904
Molecular Formula
C10H17NO3
Molecular Weight
199
No. Hydrogen Bond Acceptors
4
No. Hydrogen Bond Donors
1
Details
GC Conditions (App ID: 20809)
Column:
Part No:
Phase Name:
ZB-5MS
Injection:
Splitless @ 250°C, 1.0 μL
Oven Program:
60ºC for 1 min to 220ºC at 20ºC for 3 min to 320ºC at 30ºC/min
Carrier Gas:
Constant Flow Helium, 1.2 mL/min
Detection:
Mass Selective (MSD) (320°C)
Sample Notes:
Sample is derivatized with BSTFA at 70ºC for 30 minutes. Sim mode: Ecgonine methylester (271), Cocaine (182), Cocaethylene (196), Benzoylecgonine (240), Dihydrocodeine (373), Codeine (371), Ethylmorphine (385), Morphine (429), 6-MAM (399)

Order Items Used in This Application

Similar Applications
search
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-1, GC Cap. Column 15 m x 0.25 mm x 0.25 µm, Ea
Standard