FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Neonicotinoids by LC/MS/MS on a Kinetex Biphenyl
22781
Separation Mode: Reversed Phase
Reversed Phase
Biphenyl
HPLC
Insecticides

Applications

Neonicotinoids by LC/MS/MS on a Kinetex Biphenyl
Analytes
1Dinotefuran
2Nitenpyram
3D3-Thiamethoxam
4Thiamethoxam
5D3-Clothianidin
6Clothianidin
7D4-Imidacloprid
8Imidacloprid
9D3-Acetamiprid
10Acetamiprid
11Thiacloprid
Compound Name
Dinotefuran
CID
197701
Molecular Formula
C7H14N4O3
Molecular Weight
202
No. Chiral Centers:
1
No. Hydrogen Bond Acceptors
6
No. Hydrogen Bond Donors
2
Details
LC Conditions (App ID: 22781)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
Biphenyl
Elution Type:
Gradient
Mobile Phase:
A: 0.1% Formic acid in water
B: 0.1% Formic acid in Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
10955
233070
33.1955
46955
Flow Rate:
0.5 mL/min
Temperature:
22°C
Detection:
Tandem Mass Spec (MS-MS) (22°C)
Detector Info:
SCIEX 4500 QTRAP®

Order Items Used in This Application