For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Pyrethroids Extracted from Marine Sediment using QuEChERS by LCMSMS
23558
Separation Mode: Reversed Phase
Reversed Phase
EVO C18
HPLC
Insecticides

Applications

Pyrethroids Extracted from Marine Sediment using QuEChERS by LCMSMS
Analytes
1Cyfluthrin
2Cyhalothrin
3Cypermethrin
4Deltamethrin
5Esfenvalerate
6cis-Permethrin
7trans-Permethrin
8Bifenthrin
Compound Name
Baythroid
CID
104926
Molecular Formula
C22H18Cl2FNO3
Molecular Weight
434
No. Chiral Centers:
3
No. Hydrogen Bond Acceptors
5
Details
LC Conditions (App ID: 23558)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
EVO C18
Elution Type:
Gradient
Mobile Phase:
A: 5mM ammonium acetate
B: methanol
Gradient Profile:
Step No.Time(min)%A%B
105050
21595
33595
43.015050
565050
Flow Rate:
700 µL/min
Temperature:
22°C
Detection:
Tandem Mass Spec (MS-MS) (22°C)
Detector Info:
SCIEX Triple Quad™ 4500 System

Order Items Used in This Application

Similar Applications
search
No Data