FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Mitragynine and 7-Hydroxy (7-OH) Mitragynine from Raw Materials by HPLC-UV using Kinetex EVO 5 µ
23875
Separation Mode: Reversed Phase
Reversed Phase
EVO C18
HPLC

Applications

Mitragynine and 7-Hydroxy (7-OH) Mitragynine from Raw Materials by HPLC-UV using Kinetex EVO 5 µ
Analytes
17-OH Mitragynine
2Mitragynine

Analyte not found

Details
LC Conditions (App ID: 23875)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
EVO C18
Elution Type:
Gradient
Mobile Phase:
A: 5.0 mM Ammonium bicarbonate (pH 9.50)
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
107030
2173070
317.93070
4187030
Flow Rate:
1.5 mL/min
Temperature:
25°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 226 nm (22°C)