For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Barbiturates from urine by Strata X (60mg/3cc) (TIC)
23895
Separation Mode: Application Specific
Application Specific
ZB-Drug-1
GC
Anticonvulsants
Clinical
Forensics/Toxicology

Applications

Barbiturates from urine by Strata X (60mg/3cc) (TIC)
Analytes
1Butalbital-d5
2Butalbital
3Amobarbital
4Pentobarbital
5Secobarbital
6Phenobarbital-d5
7Phenobarbital

Analyte not found

Details
GC Conditions (App ID: 23895)
Column:
Part No:
Phase Name:
ZB-Drug-1
Injection:
Split 10:1 @ 280°C, 1.0 μL
Oven Program:
145 °C to 300 °C @ 25 °C/min
Carrier Gas:
Constant Flow Helium, 0.6 mL/min
Detection:
Mass Selective (MSD) (320°C)
Sample Notes:
Analytes are 300ng/mL in urine derivatized with TMPAH Sim ions: (Quant ion*) D5-Butalbital (*201,184,214): Butalbital (*196,181,138): Amobarbital (*169,184,225): Pentobarbital (*169,184,225): Secobarbital (*196,181,138): D5-Phenobarbital (*233,151,122): Phenobarbital (*232,146,117)
Sample Preparation Details
Description:
Strata-X 33 µm Polymeric Reversed Phase, 60 mg / 3 mL, Tubes, 50/Pk
Part Number:
Pre Treatment Note:
Reconstitute with 100uL Acetonitrile and 50uL of Thermo Scientific MethElute Reagent (TS49301)
Condition 1:
Methanol
Condition 2:
Water
Condition 3:
20mM potassium phosphate pH 2.7
Sample PreTreatment:
Reconstitute with 100uL Acetonitrile and 50uL of Thermo Scientific MethElute Reagent (TS49301)
Wash 1:
Water
Wash 2:
100mM acetic acid
Evaporation:
Evaporate to dryness
1
Condition
Methanol
2
Wash
100mM acetic acid
3
Condition
20mM potassium phosphate pH 2.7
1
Shake
85:15 Ethyl Acetate:Isopropyl Alcholol
1
Wash
Water
2
Condition
Water

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm C18 100 Å, LC Column 100 x 2.1 mm, Ea
Reversed Phase
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Kinetex 2.6 µm C18 100 Å, LC Column 100 x 2.1 mm, Ea
Reversed Phase
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-5MS, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Standard