FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
PAH with Surrogate Standards on Envirosep-PP (UV), EPA Method 610
5299
EPA 610
Carcinogens
Environmental
Pollutant / Contaminant
Environmental / Health

Applications

PAH with Surrogate Standards on Envirosep-PP (UV), EPA Method 610
Analytes
1Nitrobenzene
2Naphthalene
3Acenaphthylene
4Acenaphthene
5Fluorene
6Phenanthrene
7Anthracene
8Fluoranthene
9Pyrene
10p-Terphenyl
11Benzo[a]anthracene
12Chrysene
13Benzo[e]pyrene
14Benzo[b]fluoranthene
15Benzo[k]fluoranthene
16Benzo[a]pyrene
17Dibenz[a,h]anthracene
18Benzo[g,h,i]perylene
19Indeno[1,2,3-cd]pyrene
Compound Name
Nitrobenzene
CID
7416
Molecular Formula
C6H5NO2
Molecular Weight
123
No. Hydrogen Bond Acceptors
2
Details
Official Method: EPA 610
LC Conditions (App ID: 5299)
Elution Type:
Gradient
Mobile Phase:
A: Water
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
106040
2250100
Flow Rate:
2 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 254 nm (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282
Similar Applications
search
Kinetex 2.6 µm Polar C18 100 Å, LC Column 100 x 4.6 mm, Ea
Reversed Phase
Kinetex 5 µm EVO C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Synergi 2.5 µm Hydro-RP 100 Å, LC Column 100 x 2 mm, Ea
Reversed Phase
Synergi 2.5 µm Polar-RP 100 Å, LC Column 100 x 2 mm, Ea
Reversed Phase
Synergi 4 µm Polar-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase