For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Anti-Inflammatory/Analgesic Drug on Prodigy 5 ODS (3)
10098
Separation Mode: Reversed Phase
Reversed Phase
ODS-3
HPLC
Anti-Inflammatory Agents, Non-Steroidal
Clinical
Pharmaceutical/Biopharmaceutical
Pharmaceutical
Therapeutic / Clinical

Applications

Anti-Inflammatory/Analgesic Drug on Prodigy 5 ODS (3)
Analytes
1Butyrophenone
2Naproxen
Compound Name
Butyrophenone
CID
10315
Molecular Formula
C10H12O
Molecular Weight
148
No. Hydrogen Bond Acceptors
1
Details
LC Conditions (App ID: 10098)
Column:
Brand Name:
Part No:
Phase Name:
ODS-3
Elution Type:
Isocratic
Mobile Phase:
A: Acetonitrile
B: Water
C: Acetic acid
Gradient Profile:
Step No.Time(min)%A%B%C
1050491
Flow Rate:
1.2 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 254 nm (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Similar Applications
search
Synergi 4 µm Fusion-RP 80 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Gemini 5 µm C18 110 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Fusion-RP 80 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Gemini 5 µm C18 110 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase