FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA Method 610-QC-FL, PAH's - Florida
14482
Separation Mode: Reversed Phase
Reversed Phase
EPA 610
C18(2)
HPLC
Carcinogens
Environmental
Pollutant / Contaminant
Environmental / Health

Applications

EPA Method 610-QC-FL, PAH's - Florida
Analytes
1Naphthalene
2Acenaphthylene
31-Methylnaphthalene
42-Methylnaphthalene
5Acenaphthene
6Fluorene
7Phenanthrene
8Anthracene
9Fluoranthene
10Pyrene
11Benz[a]anthracene
12Chrysene
13Benzo[b]fluoranthene
14Benzo[k]fluoranthene
15Benzo[a]pyrene
16Dibenz[a,h]anthracene
17Benzo[g,h,i]perylene
18Indeno[1,2,3-cd]pyrene
Compound Name
Naphthalene
CID
931
Molecular Formula
C10H8
Molecular Weight
128
Details
Official Method: EPA 610
LC Conditions (App ID: 14482)
Column:
Brand Name:
Part No:
Phase Name:
C18(2)
Elution Type:
Gradient
Mobile Phase:
A: Water
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
106040
256040
3300100
Flow Rate:
1.55 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 210 nm (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Similar Applications
search
Kinetex 3.5 µm PAH, LC Column 100 x 4.6 mm, Ea
Reversed Phase
Kinetex 3.5 µm PAH, LC Column 100 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Max-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Max-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase