FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
ASTM Method D2887-97: Boiling Range Distribution of Petroleum Fractions on ZB-MR1
16255
Separation Mode: Application Specific
Application Specific
ASTM D 2887
ZB-MultiResidue-1
GC
Carcinogens
Fuels
Petrochemical
Energy / Petroleum / Plastic

Applications

ASTM Method D2887-97: Boiling Range Distribution of Petroleum Fractions on ZB-MR1
Analytes
1n-Nonane (C9)
2n-Decane (C10)
3n-Dodecane (C12)
4n-Tetradecane (C14)
5n-Hexadecane (C16)
6n-Octadecane (C18)
7n-Nonadecane (C19)
8n-Eicosane (C20)
9n-Docosane (C22)
10n-Tetracosane (C24)
11n-Hexacosane (C26)
12n-Octacosane (C28)
Compound Name
N-Nonane
CID
8141
Molecular Formula
C9H20
Molecular Weight
128
Details
Official Method: ASTM D 2887
GC Conditions (App ID: 16255)
Column:
Part No:
Phase Name:
ZB-MultiResidue-1
Injection:
Split 10:1 @ 300°C, 0.5 μL
Oven Program:
70 °C to 340 °C at 10 °C/min for 1 min.
Carrier Gas:
Constant Flow Helium, 3.0 mL/min
Detection:
Flame Ionization (FID) (350°C)
Sample Notes:
Diesel Fuel at 600 ppm in Hexane

Order Items Used in This Application

Similar Applications
search
Zebron ZB-MultiResidue-2, GC Cap. Column 30 m x 0.25 mm x 0.20 µm, Ea
Application Specific
Zebron ZB-MultiResidue-2, GC Cap. Column 30 m x 0.25 mm x 0.20 µm, Ea
Application Specific
Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific