FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA 8081 with Pentachlorophenol on ZB-MR2
16277
Separation Mode: Application Specific
Application Specific
EPA 8081A
ZB-MultiResidue-2
GC
Insecticides
Environmental
Pollutant / Contaminant
Environmental / Health

Applications

EPA 8081 with Pentachlorophenol on ZB-MR2
Analytes
1alpha-BHC
2Pentachlorophenol
3gamma-BHC
4b-BHC
5delta-BHC
6Heptachlor
7Aldrin
8Heptachlor epoxide
9trans-Chlordane
10cis-Chlordane
11Endosulfan I
124,4'-DDE
13Dieldrin
14Endrin
154,4'-DDD
16Endosulfan II
17Endrin aldehyde
184,4'-DDT
19Endosulfan sulfate
20Methoxychlor
21Endrin ketone
Compound Name
Lindane
CID
727
Molecular Formula
C6H6Cl6
Molecular Weight
290
Details
Official Method: EPA 8081A
GC Conditions (App ID: 16277)
Column:
Part No:
Phase Name:
ZB-MultiResidue-2
Injection:
On-Column @ 103°C, 1.0 μL
Oven Program:
100 °C fo r0.5 min to 220 °C at 35 °C/min to 340 °C at 20 °C/min for 2 min
Carrier Gas:
Constant Flow Helium, 3.2 mL/min
Detection:
Electron Capture (ECD) (340°C)
Sample Notes:
Analytes are 50ppb in Isooctane. A 0.5 m 0.32 mm guard was used with a 'Y' splitter to connect columns to the inlet.

Order Items Used in This Application

Similar Applications
search
Zebron ZB-SemiVolatiles, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-SemiVolatiles, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-SemiVolatiles, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-35HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature