FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Batch to Batch comparison using Water-Soluble Vitamins: Luna 3µm HILIC, pH 3.2
16438
Vitamin B Complex
Pharmaceutical
Therapeutic / Clinical

Applications

Batch to Batch comparison using Water-Soluble Vitamins: Luna 3µm HILIC, pH 3.2
Analytes
1PABA
2Nicotinamide
3Riboflavin
4Nicotinic acid
5Pyridoxine
6Thiamine
Compound Name
4-Aminobenzoic Acid
CID
978
Molecular Formula
C7H7NO2
Molecular Weight
137
No. Hydrogen Bond Acceptors
3
No. Hydrogen Bond Donors
2
Details
LC Conditions (App ID: 16438)
Elution Type:
Gradient
Mobile Phase:
A: Acetonitrile
B: 100mM Ammonium Formate pH 3.2
C: Water
Gradient Profile:
Step No.Time(min)%A%B%C
1090100
215501040
Flow Rate:
1 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Diode Array (PDA) @ 260 nm (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, HILIC 4 x 3.0 mm, 10/Pk Part No.: AJ0-8329 - Holder Part No.: KJ0-4282