FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA 552.2 on ZB-MR2 (Carrier Gas: Hydrogen)
16570
Separation Mode: Application Specific
Application Specific
EPA 552
ZB-MultiResidue-2
GC
Caustics
Environmental
Pollutant / Contaminant
Environmental / Health

Applications

EPA 552.2 on ZB-MR2 (Carrier Gas: Hydrogen)
Analytes
1Chloroacetic acid
2Bromoacetic acid
3Dichloroacetic acid
4Dalapon
5Trichloroacetic acid
6Bromochloroacetic acid
71,2,3-Trichloropropane
8Dibromoacetic acid
9Bromodichloroacetic acid
10Chlorodibromoacetic acid
112,3-Dibromopropionic acid
12Tribromoacetic acid
Compound Name
Chloroacetic Acid
CID
300
Molecular Formula
C2H3ClO2
Molecular Weight
94
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
1
Details
Official Method: EPA 552
GC Conditions (App ID: 16570)
Column:
Part No:
Phase Name:
ZB-MultiResidue-2
Injection:
Splitless @ 250°C, 1.0 μL
Oven Program:
40 °C to 80 °C @ 12 °C/min for 4 min, to 225 °C @ 30 °C/min.
Carrier Gas:
Constant Flow Hydrogen, 2.0 mL/min
Detection:
Electron Capture (ECD) (250°C)
Sample Notes:
Standard concentration mix 0.5 ppm

Order Items Used in This Application

Similar Applications
search
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-35HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-35HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-35, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Standard
Zebron ZB-MultiResidue-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-XLB, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-5MS, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-5, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-5, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-1701, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard