For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA Method 8151A on ZB-XLB (Carrier Gas: Helium)
16765
Separation Mode: Standard
Standard
EPA 8151
ZB-XLB
GC
Herbicides
Environmental

Applications

EPA Method 8151A on ZB-XLB (Carrier Gas: Helium)
Analytes
1Dalapon
23,5-Dichlorobenzoic acid
34-Nitrophenol
4DCAA
5Dicamba
6MCPP
7MCPA
8Dichloroprop
9DBOB
102,4-D
11contaminant
12Pentachlorophenol
13Silvex
14Chloramben
152,4,5-T
16Dinoseb
172,4-DB
18Bentazon
19Picloram
20Dacthal
21Acifluorfen
Compound Name
Kenapon
CID
6418
Molecular Formula
C3H4Cl2O2
Molecular Weight
142
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
1
Details
Official Method: EPA 8151
GC Conditions (App ID: 16765)
Column:
Part No:
Phase Name:
ZB-XLB
Injection:
Splitless @ 250°C, 1.0 μL
Oven Program:
50 °C for 1 min to 195 °C @ 35 °C/min for 6 min to 205 °C @ 3 °C/min for 2 min to 320 °C @ 35 °C/min for 2 min.
Carrier Gas:
Constant Flow Helium, 1.5 mL/min
Detection:
Electron Capture (ECD) (335°C)
Sample Notes:
Anlaytes are 0.5 ppm in Hexane

Order Items Used in This Application

Similar Applications
search
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-35HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-35HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature