For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Food - Quantitative Analysis of Glyphosate, Glufosinate and AMPA by LC/MS/MS (FMOC derivatization)
17960
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Herbicides
Pesticide / Insecticide
Environmental / Health

Applications

Food - Quantitative Analysis of Glyphosate, Glufosinate and AMPA by LC/MS/MS (FMOC derivatization)
Analytes
1Glyphosate-FMOC
2Glufosinate-FMOC
3AMPA-FMOC
Compound Name
Glyphosate
CID
3496
Molecular Formula
C3H8NO5P
Molecular Weight
169
No. Hydrogen Bond Acceptors
6
No. Hydrogen Bond Donors
4
Details
LC Conditions (App ID: 17960)
Column:
Brand Name:
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: Water + 2mM Ammonium Carbonate Buffer (pH=9)
B: Methanol
Gradient Profile:
Step No.Time(min)%A%B
10991
34991
4103763
5113763
625595
Flow Rate:
350 µL/min
Temperature:
30°C
Detection:
Mass Spectrometer (MS) (300°C)
Detector Info:
Glyphosate-FMOC m/z 390>168 Glufosinate-FMOC m/z 402>180 AMPA-FMOC m/z 332>110 The Application Was Supplied Courtesy of: AQUA SERVICE SCHWERIN GMBH, Herr Uwe Böland Bestimmung erfolgt nach Derivatisierung mit FMOC gemäß ISO 21458
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C18 4 x 2.0mm, 10/Pk Part No.: AJ0-7596 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Similar Applications
search
No Data