FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
TCA 10 on Luna 5 um C18(2) 150 x 4.6 on Agilent 1100 1.0 mL/min
18247
Separation Mode: Reversed Phase
Reversed Phase
C18(2)
HPLC
Antidepressive Agents, Tricyclic
Organic Chemical
Chemical Compound

Applications

TCA 10 on Luna 5 um C18(2) 150 x 4.6 on Agilent 1100 1.0 mL/min
Analytes
1Tianeptine
2Desmethyldoxepin
3Protriptyline
4Desipramine
5Imipramine
6Amoxapine
7Doxepin
8Nortriptyline
9Amitriptyline
10Clomipramine
Compound Name
Tianeptine
CID
68870
Molecular Formula
C21H25ClN2O4S
Molecular Weight
436
No. Chiral Centers:
1
No. Hydrogen Bond Acceptors
6
No. Hydrogen Bond Donors
2
Details
LC Conditions (App ID: 18247)
Column:
Brand Name:
Part No:
Phase Name:
C18(2)
Elution Type:
Isocratic
Mobile Phase:
A: 20 mM Potassium Phosphate pH 7.0
B: 50:50 MeOH and ACN
Gradient Profile:
Step No.Time(min)%A%B
104852
Flow Rate:
1 mL/min
Temperature:
40°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 254 nm (22°C)
Detector Info:
with 0.17mm ID green capillaries and needle seat, Std flow cell, detector rate @1.0s
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282