FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Calibration Curve for Melamine Extraction (20-2000ng/mL range) from Strata Melamine on Luna HILIC
18467
Pollutant / Contaminant
Environmental / Health

Applications

Calibration Curve for Melamine Extraction (20-2000ng/mL range) from Strata Melamine on Luna HILIC
Analytes
1Melamine
2Melamine 13C3-Amino 15N3
Compound Name
Melamine
CID
7955
Molecular Formula
C3H6N6
Molecular Weight
126
No. Hydrogen Bond Acceptors
6
No. Hydrogen Bond Donors
3
Details
LC Conditions (App ID: 18467)
Elution Type:
Isocratic
Mobile Phase:
A: 90/10 ACN/100 mM Ammonium Formate, pH 3.2
Gradient Profile:
Step No.Time(min)%A
13.5100
Flow Rate:
0.4 mL/min
Temperature:
22°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
* MS in -ve ionization till 1.7 mins then to +ve till 3.5 mins
Sample Note:
LC/MS/MS Condition: Column: Luna 3µm HILIC, 100x2.0mm; Flow rate: 0.4ml/min Mobile Phase: Isocratic; 90/10 ACN/100mM NH4 Formate pH 3.2; MS Detector: API 3000 LC/MS/MS with ESI (TurboIonSpray®) (www.Appliedbiosystems.com), Switch from -ve Ion Mode (for cyanuric) to Positive Ion (for Melamine); MRM; heater gas flow 7000 cc/min; heater temperature 450 °C Melamine QC - 200 ng % Accuracy 99.45% % CV 6% Calibration Curve of Extracted Samples conc. (ng) % Accuracy 20 102% 40 96% 200 104% 400 98% 2000 100.03%