FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA Methods 508 and 8081A by ECD on ZB-35HT
18768
Separation Mode: High Temperature
High Temperature
EPA 508
ZB-35HT
GC
Insecticides
Environmental

Applications

EPA Methods 508 and 8081A by ECD on ZB-35HT
Analytes
1DBCP
2Hexachlorocyclopentadiene
3Etridiazole
4Chloroneb
5Propachlor
6Di-allate
7Hexachlorobenzene
8alpha-BHC
9PCNB
10gamma-BHC
11b-BHC
12Alachlor
13Heptachlor
14Dichlone
15delta-BHC
16Chlorothalonil
17Aldrin
18Dacthal
19Heptachlor epoxide
20trans-Chlordane
21trans-Nonachlor
22cis-Chlordane
23Endosulfan I
244,4'-DDE
25Dieldrin
26Chloropropylate
27Chlorobenzilate
28Endrin
29Nitrofen
30Kepone
314,4'-DDD
32Endosulfan II
334,4'-DDT
34Endrin aldehyde
35Endosulfan sulfate
36Captafol
37Methoxychlor
38Endrin ketone
39Permethrin
40Permethrin
Compound Name
Fumazone
CID
7280
Molecular Formula
C3H5Br2Cl
Molecular Weight
236
No. Chiral Centers:
1
Details
Official Method: EPA 508
GC Conditions (App ID: 18768)
Column:
Part No:
Phase Name:
ZB-35HT
Injection:
Split 20:1 @ 200°C, 1.0 μL
Oven Program:
90 °C for 0.5 min to 210 °C @ 30 °C/min. for 0.5 min to 230 °C @ 10 °C/min for 2 min to 260 °C @ 5 °C/min for 2 min to 300 °C @ 6 °C/min for 2 min
Carrier Gas:
Constant Flow Helium, 0.6 mL/min
Detection:
Electron Capture (ECD) (350°C)
Sample Notes:
All analytes were 10 ppm in iso-octane

Order Items Used in This Application

Similar Applications
search
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-35HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-MultiResidue-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.32 mm x 0.50 µm, Ea
Application Specific