FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA Method 8151 - Chlorinated Herbicides on a ZB-35-HT(Dual column)
18788
Separation Mode: High Temperature
High Temperature
EPA 8081A
ZB-35HT
GC
Herbicides
Environmental
Pesticide / Insecticide
Environmental / Health

Applications

EPA Method 8151 - Chlorinated Herbicides on a ZB-35-HT(Dual column)
Analytes
1Dalapon
23,5-Dichlorobenzoic acid
34-Nitrophenol
4DCAA
5Dicamba
6MCPP
7MCPA
8DBOB
9Dichlorprop
102,4-D
11Pentachlorophenol
12contaminant
13Silvex
14Chloramben
152,4,5-T
16Dinoseb
172,4-DB
18Bentazon
19Picloram
20Dacthal
21Acifluorfen
Compound Name
Kenapon
CID
6418
Molecular Formula
C3H4Cl2O2
Molecular Weight
142
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
1
Details
Official Method: EPA 8081A
GC Conditions (App ID: 18788)
Column:
Part No:
Phase Name:
ZB-35HT
Injection:
Splitless 0:1 @ 250°C, 1.0 μL
Oven Program:
50 °C for 0.5 min to 100 °C @ 25 °C/min to 320 °C @ 12 °C/min for 2 min
Carrier Gas:
Constant Flow Helium, 2.4 mL/min
Detection:
Electron Capture (ECD) (300°C)
Sample Notes:
Analytes are 50ppb in isooctane. Dual column with ZB-XLB-HT

Order Items Used in This Application

Similar Applications
search
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-35HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-35HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature