For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Clarity OTX- Recovery Data- MS purity data of 27mer P-S peak from 18868
18869
Separation Mode: Reversed Phase
Reversed Phase
Oligo-RP
HPLC
Life Science
Pharmaceutical/Biopharmaceutical
Oligonucleotide
Biochemical Compound / Nutrient

Applications

Clarity OTX- Recovery Data- MS purity data of 27mer P-S peak from 18868
Analytes
127 mer P-S Oligonucleotide

Analyte not found

Details
LC Conditions (App ID: 18869)
Column:
Brand Name:
Part No:
Phase Name:
Oligo-RP
Elution Type:
Gradient
Mobile Phase:
A: 200mM HFIP/ 14mM TEA/ 2% MeOH/ water
B: 200mM HFIP/ 14mM TEA/ 98% MeOH/ water
Gradient Profile:
Step No.Time(min)%A%B
109010
258515
3203565
420.1595
Flow Rate:
0.25 mL/min
Temperature:
40°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 260 nm (22°C)
Detector Info:
OD260 0.159, 12ug/ml 97% recovery
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Clarity Oligo-RP 4 x 2.0mm ID, 10/Pk Part No.: AJ0-8134 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Similar Applications
search
Clarity 3 µm Oligo-RP, LC Column 150 x 2 mm, Ea
Reversed Phase
Clarity 3 µm Oligo-RP, LC Column 150 x 2 mm, Ea
Reversed Phase