For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Hu IgG2k aggregates on BioSep3000
18906
Separation Mode: Gel Filtration Chromatography (GFC)
Gel Filtration Chromatography (GFC)
SEC-s3000
HPLC
Pharmaceutical/Biopharmaceutical
Life Science
Protein
Biochemical Compound / Nutrient

Applications

Hu IgG2k aggregates on BioSep3000
Analytes
1IgG
2IgG2 trimer
3IgG dimer
4IgG dimer
5IgG monomer
6IgG

Analyte not found

Details
LC Conditions (App ID: 18906)
Column:
Brand Name:
Part No:
Phase Name:
SEC-s3000
Elution Type:
Isocratic
Mobile Phase:
A: 100mM Sodium Phosphate pH 6.8
Gradient Profile:
Step No.Time(min)%A
10100
Flow Rate:
1 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 220 nm (22°C)
Sample Note:
Application Focus: An example of Ig-G aggregate testing on BioSep 3000 Analyzing Ig-G based therapeutics for aggregates is probably the most common application that is performed on a GFC column. Most users are attempting to get best resolution possible between the Ig-G monomer peak and any associated dimers in the sample. In this example human Ig-G2 is injected on a BioSep 3000 column (300x7.8 mm) and dimer and aggregate is quantitated. Results show that the dimer peak of Ig-G is well resolved from the monomer peak, and that there appears to be two different dimer forms that are partially resolved by the BioSep 3000 column. In addition, there appears to be aggregate at the total excluded void of the column and the appearance of a possible trimer peak ahead of the dimer peak. Finally, there is a fragment peak that elutes after the Ig-G monomer peak, most likely attributed to an Ig-G that is missing one of its Fab fragment. These results show the utility of using BioSep 3000 for antibody aggregate analysis.

Order Items Used in This Application

Similar Applications
search
Aeris 3.6 µm WIDEPORE XB-C8 200 Å, LC Column 100 x 2.1 mm, Ea
Reversed Phase
Aeris 3.6 µm WIDEPORE XB-C8 200 Å, LC Column 100 x 2.1 mm, Ea
Reversed Phase
Aeris 3.6 µm WIDEPORE XB-C8 200 Å, LC Column 100 x 2.1 mm, Ea
Reversed Phase
Aeris 3.6 µm WIDEPORE XB-C8 200 Å, LC Column 100 x 2.1 mm, Ea
Reversed Phase
Aeris 3.6 µm WIDEPORE XB-C18 200 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Aeris 3.6 µm PEPTIDE XB-C18 100 Å, LC Column 150 x 2.1 mm, Ea
Reversed Phase
Aeris 3.6 µm PEPTIDE XB-C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase