For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA 552.2 by Strata-SAX on a ZB-MR2
19287
Separation Mode: Application Specific
Application Specific
EPA 552
ZB-MultiResidue-2
GC
Caustics
Pollutant / Contaminant
Environmental / Health

Applications

EPA 552.2 by Strata-SAX on a ZB-MR2
Analytes
1Chloroacetic acid
2Bromoacetic acid
3Dichloroacetic acid
4Dalapon
5Trichloroacetic acid
6Bromochloroacetic acid
71,2,3-Trichloropropane
8Dibromoacetic acid
9Bromodichloroacetic acid
10Chlorodibromoacetic acid
112,3-Dibromopropionic acid
12Tribromoacetic acid
Compound Name
Chloroacetic Acid
CID
300
Molecular Formula
C2H3ClO2
Molecular Weight
94
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
1
Details
Official Method: EPA 552
GC Conditions (App ID: 19287)
Column:
Part No:
Phase Name:
ZB-MultiResidue-2
Injection:
Pulsed @ 250°C, 1.0 μL
Oven Program:
30°C for 4 min to 50°C @ 30°C/min for 1 min to 70°C @ 15°C/min for 1 min to 115°C @ 20°C/min for 2 min to 200°C @ 30°C/min.
Carrier Gas:
Constant Flow Helium, 2.3 mL/min
Detection:
Electron Capture (ECD) (340°C)
Sample Notes:
Derivatization: Add 1mL of MTBE to the eluent in a screw cap vial or test tube. Heat the sample at 50ºC for 2 hours. Remove and allow the sample to cool to room temperature. Slowly add 10mL of saturated sodium bicarbonate and invert the sample several times releasing carbon dioxide gas frequently. Aspirate 100µL of the organic layer and transfer to a 2mL vial with insert. Add 10µL of internal standard (1,2,3-Trichloropropane) and inject into GC-ECD system.