FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
USP Glycerin Purity Assay on ZB-BAC-2
19291
Separation Mode: Application Specific
Application Specific
ZB-BAC-2
GC
Cryoprotective Agents
Pharmaceutical/Biopharmaceutical
Pollutant / Contaminant
Environmental / Health

Applications

USP Glycerin Purity Assay on ZB-BAC-2
Analytes
1Ethylene glycol
22,2,2-Trichloroethanol
3Diethylene glycol
4Glycerin
Compound Name
Ethylene Glycol
CID
174
Molecular Formula
C2H6O2
Molecular Weight
62
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
2
Details
GC Conditions (App ID: 19291)
Column:
Part No:
Phase Name:
ZB-BAC-2
Injection:
Split 10:1 @ 220°C, 1.0 μL
Oven Program:
100 °C for 4 min to 120 °C @ 50 °C/min for 10 min to 220 °C @ 50 °C/min for 6 min
Carrier Gas:
Constant Flow Helium, 4.5 mL/min
Detection:
Flame Ionization (FID) (250°C)
Sample Notes:
Concentrations were 2.0 mg/mL Glycerin, 0.050 mg/mL ethylene glycol, 0.050 mg/mL diethylene glycol, and 0.10 mg/mL 2,2,2-trichloroethanol (IS) Run as a dual column analysis with ZB-BAC-1 Resolution = 11.09 between glycerin and diethylene glycol