For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Bivalirudin Purification-Fraction 28.8 min
19459
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Anti-Inflammatory Agents, Non-Steroidal

Applications

Bivalirudin Purification-Fraction 28.8 min
Analytes
1Front impurities
2Bivalirudin

Analyte not found

Details
LC Conditions (App ID: 19459)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 0.1% TFA in Water
B: 0.1% TFA in ACN
Gradient Profile:
Step No.Time(min)%A%B
107723
286931
38.17723
4117723
Flow Rate:
1.85 mL/min
Temperature:
0°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 220 nm (22°C)

Order Items Used in This Application

Similar Applications
search
No Data