For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Humic Acids using Kinetex C8 50x2.1 mm on ABI LC-MS 4000
19662
Separation Mode: Reversed Phase
Reversed Phase
C8
HPLC
Environmental

Applications

Humic Acids using Kinetex C8 50x2.1 mm on ABI LC-MS 4000
Analytes
1Humic acids

Analyte not found

Details
LC Conditions (App ID: 19662)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C8
Elution Type:
Gradient
Mobile Phase:
A: 5mM Ammoniium Acetate
B: Methanol
Gradient Profile:
Step No.Time(min)%A%B
108515
22595
36595
46.018515
Flow Rate:
0.4 mL/min
Temperature:
0°C
Detection:
Mass Spectrometer (MS) @ 580.4 amu (22°C)
Detector Info:
SCIEX API 4000™
Sample Preparation Details
Description:
Strata PAH, 1.5 g / 6 mL, Tubes, 30/Pk
Part Number:
Condition 1:
Dichloromethane
Condition 2:
Methanol
Condition 3:
MilliQ Water
Wash 1:
50:50 Methanol:MilliQ Water
Evaporation:
Evaporate to dryness
1
Condition
Dichloromethane
2
Condition
Methanol
3
Condition
MilliQ Water
1
Shake
Dichloromethane
1
Condition
Dichloromethane

Order Items Used in This Application

Similar Applications
search
No Data