For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
LC/MS/MS Analysis of Propoxyphene from urine using Strata-X-Drug B on a Kinetex C18 2.6µm, 50x2.1mm
19704
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Analgesics, Opioid
Forensics/Toxicology

Applications

LC/MS/MS Analysis of Propoxyphene from urine using Strata-X-Drug B on a Kinetex C18 2.6µm, 50x2.1mm
Analytes
1Propoxyphene-d11
2Propoxyphene
Compound Name
Levopropoxyphene
CID
15330
Molecular Formula
C22H29NO2
Molecular Weight
339
No. Chiral Centers:
2
No. Hydrogen Bond Acceptors
3
Details
LC Conditions (App ID: 19704)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 0.1% Formic acid
B: 0.1% Formic acid in acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
106040
23.50100
33.516040
46.56040
Flow Rate:
400 µL/min
Temperature:
0°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
SCIEX API 3000™ Source: Cur-11, CAD-12, IS-4500, Tem-450, NEB-12 Dwell: 25ms Polarity: Positive
Sample Preparation Details
Description:
Strata-X-Drug B 33 µm Polymeric Strong Cation, 60 mg / 6 mL, Tubes, 30/Pk
Part Number:
Pre Treatment Note:
Evaporate to dryness under a stream of nitrogen at 50 C . Reconstitute in 1 mL of 40% acetonitrile.
Sample PreTreatment:
Evaporate to dryness under a stream of nitrogen at 50 C . Reconstitute in 1 mL of 40% acetonitrile.
Wash 1:
2 mLs of 100mM Sodium acetate (pH 5.0) buffer
Wash 2:
2 mLs of Methanol
Evaporation:
Evaporate to dryness
1
Shake
2mLs of Ethyl acetate:Isopropanol:NH4OH (7:2:1)
1
Wash
2 mLs of 100mM Sodium acetate (pH 5.0) buffer
2
Wash
2 mLs of Methanol

Order Items Used in This Application

Similar Applications
search
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-5MS, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Standard