FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
LC/MS/MS of Extended Opiates from urine using Strata-X-Drug B on a Kinetex C18 2.6µm, 50x2.1mm
19705
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Adjuvants
Forensics/Toxicology
Clinical

Applications

LC/MS/MS of Extended Opiates from urine using Strata-X-Drug B on a Kinetex C18 2.6µm, 50x2.1mm
Analytes
1Oxymorphone-d3
2Oxymorphone
3Morphine-d3
4Morphine
5Hydromorphone-d6
6O-Desmethyltramadol
7Hydromorphone
8Oxycodone-d6
9Oxycodone
10Codeine-d6
11Codeine
12N-Desmethyltramadol
13cis-Tramadol
14Hydrocodone-d6
15Hydrocodone
16Normeperidine-d4
17Normeperidine
18Meperidine-d4
19Meperidine
Compound Name
Oxymorphone
CID
5284604
Molecular Formula
C17H19NO4
Molecular Weight
301
No. Hydrogen Bond Acceptors
5
No. Hydrogen Bond Donors
2
Details
LC Conditions (App ID: 19705)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 5mM Ammonium acetate
B: 5mM Ammonium acetate in methanol
Gradient Profile:
Step No.Time(min)%A%B
108515
25.50100
36.50100
46.518515
598515
Flow Rate:
500 µL/min
Temperature:
0°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
SCIEX API 4000™ Source: Cur-25, CAD-7, IS-5500, Tem-600, GS1 60, GS2 55 Dwell: 20ms Polarity: Positive API 4000
Sample Preparation Details
Description:
Strata-X-Drug B 33 µm Polymeric Strong Cation, 60 mg / 6 mL, Tubes, 30/Pk
Part Number:
Pre Treatment Note:
Evaporate to dryness under a stream of nitrogen at 50 C . Reconstitute in 1 mL of 15% methanol.
Sample PreTreatment:
Evaporate to dryness under a stream of nitrogen at 50 C . Reconstitute in 1 mL of 15% methanol.
Wash 1:
2mLs of 100mM Sodium acetate (pH 5.0) buffer
Wash 2:
2mLs of Methanol
Evaporation:
Evaporate to dryness
1
Shake
2mLs of Ethyl acetate:Isopropanol:NH4OH (7:2:1)
1
Wash
2mLs of 100mM Sodium acetate (pH 5.0) buffer
2
Wash
2mLs of Methanol

Order Items Used in This Application

Similar Applications
search
No Data