FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Separation of Testosterone and Epitestosterone by LC/MS/MS using Strata-X-A and Kinetex 1.7 µm C18
19761
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Androgens
Clinical

Applications

Separation of Testosterone and Epitestosterone by LC/MS/MS using Strata-X-A and Kinetex 1.7 µm C18
Analytes
1Testosterone
2Testosterone-d3
3Epitestosterone
Compound Name
Testosterone
CID
6013
Molecular Formula
C19H28O2
Molecular Weight
288
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 19761)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 0.1% Formic Acid +1 mM Amm Formate in Water
B: 0.1% Formic Acid +1 mM Amm Formate in ACN
Gradient Profile:
Step No.Time(min)%A%B
109010
22.51090
33.51090
43.69010
Flow Rate:
0.4 mL/min
Temperature:
0°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
SCIEX API 5000™
Sample Preparation Details
Description:
Strata-X-A 33 µm Polymeric Strong Anion, 30 mg / 3 mL, Tubes, 50/Pk
Part Number:
Pre Treatment Note:
Following evaporation of elution solvent @ 50-55 C under gentle nitrogen stream; Add 50 uL 25% hydroxylamine solution and heat at 60-65 C for 5-10 min, then add 200 uL 5% aqueous formic acid and vortex the tubes Transfer the solution to autosampler vials and inject 25 uL on column
Condition 1:
Methanol
Condition 2:
Water
Sample PreTreatment:
Following evaporation of elution solvent @ 50-55 C under gentle nitrogen stream; Add 50 uL 25% hydroxylamine solution and heat at 60-65 C for 5-10 min, then add 200 uL 5% aqueous formic acid and vortex the tubes Transfer the solution to autosampler vials and inject 25 uL on column
Wash 1:
50% Methanol in Water
Evaporation:
Evaporate to dryness
1
Condition
Methanol
2
Shake
Methanol
1
Condition
Methanol
2
Shake
Methanol
1
Condition
Methanol

Order Items Used in This Application

Similar Applications
search
Kinetex 5 µm EVO C18 100 Å, LC Column 100 x 3.0 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Kinetex 1.7 µm C18 100 Å, LC Column 30 x 2.1 mm, Ea
Reversed Phase
Kinetex 1.7 µm C18 100 Å, LC Column 30 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Gemini 3 µm NX-C18 110 Å, LC Column 50 x 2 mm, Ea
Reversed Phase
Gemini 3 µm NX-C18 110 Å, LC Column 50 x 2 mm, Ea
Reversed Phase
Gemini 5 µm C18 110 Å, LC Column 50 x 2 mm, Ea
Reversed Phase
Gemini 5 µm C18 110 Å, LC Column 50 x 2 mm, Ea
Reversed Phase
Luna 5 µm CN 100 Å, LC Column 150 x 4.6 mm, Ea
Normal Phase