For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Kinetex 1.7u XB-C18 30x2.1mm: Peak capacity test on Shimadzu Nexera
19803
Separation Mode: Reversed Phase
Reversed Phase
XB-C18
HPLC
Preservatives, Pharmaceutical

Applications

Kinetex 1.7u XB-C18 30x2.1mm: Peak capacity test on Shimadzu Nexera
Analytes
1Uracil
2Methyl paraben
3Acetophenone
4Propyl paraben
5Butyl paraben
6Naphthalene
Compound Name
Uracil
CID
1174
Molecular Formula
C4H4N2O2
Molecular Weight
112
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
2
Details
LC Conditions (App ID: 19803)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
XB-C18
Elution Type:
Gradient
Mobile Phase:
A: Water
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
10982
20.4298
30.53298
Flow Rate:
3 mL/min
Temperature:
0°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 254 nm (22°C)
Detector Info:
Micro cell, 100Hz, no filter

Order Items Used in This Application

Similar Applications
search
No Data