For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
USP Simvastatin Lovastatin on Kinetex 2.6um C18 100x2.10mm using Shimadzu Nexera
19961
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Antilipemic Agents
Pharmaceutical/Biopharmaceutical

Applications

USP Simvastatin Lovastatin on Kinetex 2.6um C18 100x2.10mm using Shimadzu Nexera
Analytes
1Lovastatin
2Simvastatin
Compound Name
Lovastatin
CID
53232
Molecular Formula
C24H36O5
Molecular Weight
404
No. Hydrogen Bond Acceptors
5
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 19961)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Isocratic
Mobile Phase:
A: 32mM Sodium Phosphate pH 4.5
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
153565
Flow Rate:
0.3 mL/min
Temperature:
0°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 238 nm (22°C)